1. Background Theory
Table B1. Thermal emf in absolute mV for T-Type thermocouple with ref. =0°C. [Holman]
Temperature (°C) | Voltage (mV) |
-184.4 | -5.341 |
-156.7 | -4.745 |
-128.9 | -4.419 |
-101.1 | -3.365 |
-73.3 | -2.581 |
-45.55 | -1.626 |
-17.8 | -0.674 |
10 | 0.422 |
37.8 | 1.518 |
65.6 | 2.743 |
93.3 | 3.967 |
121.1 | 5.307 |
Figure B1. Temperature plot as a funcion of voltage displayed.
From Figure B1., we obtain the calibrition equation:
[Eqn. 1]
T: Temperature measured.
V: Voltage displayed
When a temperature measurement is not made under steady-state conditions, the energy balance for the changing thermal system may be written as:
[Eqn. 3]
.h: Convection heating transfer coefficient between the thermometer and the fluid.
A: Surface area of the thermal system.
: Temperature of the environment.
T: Temperature measured.
Eqn. 3 may also be written as the temperature of the thermometer as a function of time:
[Eqn. 4]
: Temperature of the thermometer at time zero.
Other variables are defines as in above.
Eqn. 4 is often rewritten as:
[Eqn. 5]
in which , the time constant, is defined as mc/hA
It is possible to linearize Eqn. 5 into the following form:
= [Eqn. 6]
If Eqn. 6 is plot as a function of time, the slope will be equal to , the time constant can thus be calculated from known temperature at different time intervals.
No comments:
Post a Comment